"X-Ray Microanalysis of Uranium Dioxide Particles by Using Low Electron Beam Energy"статья
Информация о цитировании статьи получена из
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 9 июня 2015 г.
Аннотация:The elemental analysis results of the various size uranium dioxide fragments, performed with a scanning electron microscope (SEM) equipped by X-ray energy dispersive analyzer (INCA xact, “Oxford Instruments”, Great Britain), are presented. The elemental composition calculation was carried out applying the preinstalled program based on the XPP matrix correction algorithm. It is shown that for the bulky UO2 samples, measured with the varied accelerating voltage in range of 6–20 kV, the uranium concentrations slightly differ (relative uncertainties doesn’t exceed ±2%). The composition of uranium dioxide particles
sized up to 0.5 μm can be defined using beam energy of 6 keV with typical for bulky sample accuracy.