Spectrophotometric and thermal lens determination of aluminum with 3-sulfo-5-nitro-4'-diethylamino-2,2'-dihydroxyazobenzeneстатья
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.
Аннотация:The conditions of the spectrophotometric and thermal lens determination of aluminum with sulfo-5-nitro-4'-diethylamino-2,2'-dihydroxyazobenzene have been compared. The limit of spectrophotometric detection of aluminum in aqueous solutions has been found to be 8 ng/mL. On the basis of the conditions of spectrophotometric determination, the conditions for thermal lens determination have been proposed (532.0 nm, exciting radiation power of 42 mW); they provide a decrease of the detection limit down to 0.6 ng/mL and an increase of the sensitivity coefficient by an order of magnitude. It has been shown that, in the case of the thermal lens determination of aluminum in water-organic mixtures (50 vol % of dimethyl sulfoxide or 30 vol % acetonitrile), the sensitivity coefficient is respectively 9.1 and 6.3-fold higher as compared with the thermal lens determination in water. As a result, the detection limits are reduced 2.5 and 10-fold, respectively. Aluminum has been determined by thermal lens spectrophotometry in Moscow's tap water using the standard addition method, its concentration being 0.79 +/- 0.07 mg/L, which is above the threshold limit value of the aluminum content of drinking water.