Effects of dead time and after-pulses in photon detector on measured statistics of stochastic radiationстатья
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 20 мая 2014 г.
Аннотация:Many physical experiments require analysis of the statistics of fluctuating radiation. In the case of an ideal single-photon detector, the contribution of photon noise to the statistics of the registered signal has been thoroughly examined. However, practical photon counters have a dead time, leading to miscounting of certain true events, and sometimes the counters generate false after-pulses.
This study investigate the impact of these two effects, and it presents the theoretical relations between the statistical moments of the radiation and the registered counts while also accounting for dead time and the probability of after-pulses. Expressions for statistical moments of any order are obtained on the basis of the generalized Poisson distribution (GPD). For counters with paralyzable dead time, alternative relations for the mean and variance are derived using generally accepted formulas.
As an example, the measurements of stellar scintillation and the result of simple experiment are considered. The results of the experimental verification of the theoretical expression confirm the need to account for the non-ideal nature of detectors in almost all similar measurements.