Analytical potentialities of an AI-05 instrument for capillary electrophoresis in the determination of metal cationsстатья

Информация о цитировании статьи получена из Scopus, Web of Science
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.

Работа с статьей


[1] Krokhin O. V., Obrezkov O. N., Shpigun O. A. Analytical potentialities of an ai-05 instrument for capillary electrophoresis in the determination of metal cations // Journal of Analytical Chemistry. — 1998. — Vol. 53, no. 4. — P. 372–374. The analytical potentialities of an AI-05 instrument for capillary zone electrophoresis were examined using the determination of alkali, alkaline-earth, and transition metal cations as an example. The reproducibility of retention times and peak areas was 0.06-0.08 and 0.1, respectively. The plot of the analytical signal of Sr2+ against the concentration in the sample was linear within the range 1 × 10-5-1 × 10-4 M. The phenomenon of discrimination of the ionic composition of a sample in the electromigration method of sample injection is discussed. The time taken to separate a mixture of K+, Na+, Mg2+, Sr2+, Ca2+, Cd2+, Co2+, Ni2+, and Zn2+ was 16 min. © 1998 MAEe cyrillic signK Hayκa/Interperiodica Publishing.

Публикация в формате сохранить в файл сохранить в файл сохранить в файл сохранить в файл сохранить в файл сохранить в файл скрыть