Measurement of Short Time Interval for the System «Laser-Time Position Sensitive Detector of Radiation"статья
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 6 мая 2016 г.
Аннотация:For substantial improvement of resolution of system «Laser Position Sensitive Detector of Radiation» the optimum method of measurement of a time interval is chosen and practical realisation of the measuring instrument which is based on microcircuit TDC-GPX is carried out. Test measurements have shown that the time interval measuring instrument has resolution 10 ps at a range of measurements to 40 ms.