Processible polyaniline as an advanced potentiometric pH transducer. Application to biosensorsстатья
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.
Аннотация:An advanced potentiometric pH transducer based on processible polyaniline (PCPAn) is reported on. Both glassy carbon and screen-printed carbon electrodes modified with PCPAn by dip-coating exhibited a fully reversible potentiometric response of approximately 90 mV/pH unit over the range from pH 3 to 9. Such a significantly higher potentiometric response of PCPAn-modified electrodes as compared with those of existing devices is explained on the basis of the thermodynamics of polyaniline redox reactions. The PCPAn-based pH transducers exhibit both good operational stability and prolonged shelf life and display a negligible response toward singly charged cations. The new thick-film pH transducer was employed for designing a potentiometric biosensor for urea. Zn the model solution which mimics blood serum, the urea-sensitive electrode has a detection limit of 10(-5) M urea and a maximum response of similar to 120 mV, The attractive performance characteristics are advantageous over those of existing pH sensors and offer great promise for sensing and biosensing applications.