Amperometric biosensor for glutamate using Prussian Blue-based "artificial peroxldase" as a transducer for hydrogen peroxideстатья
Статья опубликована в высокорейтинговом журнале
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.
Аннотация:The specially deposited Prussian Blue denoted as "artificial peroxidase" was used as a transducer for hydrogen peroxide, The electrocatalyst as stable, highly active, and selective to hydrogen peroxide reduction in the presence of oxygen, which allowed sensing of H2O2 around 0.0 V (Ag/AgCl). Glutamate oxidase was immobilized on the surface of the Prussian Blue-modified electrode in a Nafion layer using a nonaqueous enzymology approach. The calibration range for glutamate in flow injection system was 1 x 10(-7)-1 x 10(-4) M. The lowest concentration of glutamate detected (1 x 10(-7) M) and the highest sensitivity in the linear range of 0.21 A M-1 cm(-2) were achieved. The influence of reductants was practically avoided using the low potential of an indicator electrode (0.0 V Ag/AgCl). The attractive performance characteristics of the glutamate biosensor illustrate the advantages of Prussian Blue-based "artificial peroxidase" as transducer for hydrogen peroxide detection.