A new approach for determination of neuropathy target esterase activityстатья
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.
Аннотация:Neuropathy target esterase (NTE) was shown to be an excellent biochemical marker for screening of organophosphates (OPs) with respect to their ability to result in organophosphate induced delayed neurotoxicity (OPIDN). This paper describes a new biosensor approach to the analysis of NTE and its inhibitors. The method is based on the combination of NTE enzymatic hydrolysis of phenyl valerate (PV) with phenol detection by the Clark-type oxygen electrode modified by immobilized tyrosinase. The validity of this biosensor method is confirmed by the facts that the calibration curves for NTE obtained by colorimetric and flow-through electrochemical methods were nearly identical and the titration of NTE by test inhibitor mipafox was shown to yield the same pI50 values. The developed electrochemical methods can be considered as a promising approach both for serial express NTE analysis and for kinetic characteristics of NTE.